| The Barben Analyzer DO system combines our easy-maintenance 546 DO sensor with the outstanding Stratos 2402DO Analyzer / Controller, packaged into a NEMA 4X box along with a built-in air compressor to operate the included jet-cleaner. The system is available in single or dual channel, with an optional easy-mount railing attachment system. |
|
|
Sensor limits and characteristics: Measurement Ranges 0.0 to 50 ppm using 4 mil membrane 0.0 to 20 ppm using 1 mil membrane Response Time 90% within 60 sec Repeatability +/- 1% Measurement Principle Galvanic Clark electrode Sample Flow Requirements 0.4 ft/sec Membrane Thickness 1 mil & 4 mil Maximum Sensor to Analyzer Distance 750 ft Average replacement interval 1 year Shelf life Indefinite
Wetted Materials Include: CPVC, Teflon, ABS, EPR O-rings
Maximum Pressures & Temperature Ratings: Temperature: 0 to 60 C Pressure: 0 to 150 psig |